Click here to close Hello! We notice that you are using Internet Explorer, which is not supported by Xenbase and may cause the site to display incorrectly. We suggest using a current version of Chrome, FireFox, or Safari.
XB-ART-1701
Opt Lett 2005 Jun 01;3011:1351-3. doi: 10.1364/ol.30.001351.
Show Gene links Show Anatomy links

Stroboscopic ultrahigh-resolution full-field optical coherence tomography.

Moneron G , Boccara AC , Dubois A .


???displayArticle.abstract???
We present a new technique that produces en face tomographic images with a 10-micros acquisition time per image. The setup consists of an interference microscope with stroboscopic illumination provided by a xenon arc flash lamp (10-micros flashes at 15 Hz). The tomographic images are obtained from two phase-opposed interferometric images recorded simultaneously by two synchronized CCD cameras. Transverse resolution better than 1.0 microm is achieved by use of high-numerical-aperture microscope objectives. The short coherence length of the source yields an axial resolution of 0.9 microm. 3 x 3 pixel binning leads to a detection sensitivity of 71 dB. Our system is suitable for various applications, particularly in biology for in vivo cellular-level imaging.

???displayArticle.pubmedLink??? 15981530
???displayArticle.link??? Opt Lett


Species referenced: Xenopus laevis
Genes referenced: runx2